STERRAD 100NX Cassette (Used to provide hydrogen peroxide sterilant used in the STERRAD NX Steriliser) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00811-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-08-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Sterrad 100nx system cassettes are designed with a leak indicator on the outer package to provide a visual colour indication in the event hydrogen peroxide has leaked from the cassettes. the manufacturer has determined it does not have adequate data to support the effectiveness of the packaging's leak indicator for the entire duration of the sterrad 100nx system cassette's 15 month shelf life.
  • 조치
    Johnson & Johnson Medical (JJM) is requesting end users to wear additional safety equipment (protective gloves and eye wear) in the event of handling product with an expired leak indicator as a precautionary measure for devices with an expiration date prior to 1 November 2014. End users are also requested to check for signs of indicator colour change or liquid leakage prior to opening the package. If the white indicator has turned red or a liquid leak is observed, customers are asked to dispose of the cassette immediately. Any indication of leakage should also be notified to JJM.

Device

  • 모델명 / 제조번호(시리얼번호)
    STERRAD 100NX Cassette (Used to provide hydrogen peroxide sterilant used in the STERRAD NX Steriliser)Part Number: 10144Expiration dates before November 1, 2014
  • Manufacturer

Manufacturer