Stratus CS Acute Care Diagnostics System. An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00867-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-06-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare has identified a software defect in the stratus cs acute care diagnostics system regarding testpak calibration and quality control (qc) status when a testpak is due to expire within 48 hours. depending on the calibration status, either an above assay range (aar) or an inaccurate value could potentially be reported when running a testpak that is due to expire within 48 hours. the probability of occurrence is extremely unlikely and other factors such as previous results, patient presentation, and other diagnostic testing would initiate clinical questioning and reduce the potential for injury.
  • 조치
    Siemens Healthcare is advising users of the following: 1. Do not use TestPaks that are due to expire within 48 hours, OR 2. Check Reagent Status to ensure all TestPaks in use have valid calibration expiry dates beyond the TestPak lot expiry date, AND 3. If the laboratory performs routine QC and if using a TestPak that is due to expire within 48 hours, ensure that scheduled QC is performed before running a test. Siemens Healthcare is currently developing software updates to address this issue and will be providing new information as it becomes available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Stratus CS Acute Care Diagnostics System. An in vitro diagnostic medical device (IVD).Siemens Material Numbers: 10444834, 10453531, 10453570, 10453571, 10470384All software versions affectedAll serial numbers affectedARTG Number: 174699
  • 의료기기 분류등급
  • Manufacturer

Manufacturer