Surveyor S12 and Surveyor S19 Patient MonitorsMultiple part numbers and serial numbers affected 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Mortara Instrument Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01108-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-10-17
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Mortara instrument, inc. has recently become aware of a potential safety hazard involving surveyor s12 and surveyor s19 patient monitors. a hazardous situation may be created when mounting these monitors using the 75mm vesa mounting hole pattern provided in the battery cover of these monitors. if the mounting screws provided by mortara are not used and if the mounting screws used are longer than the provided screws, it is possible for the longer screws to be driven completely through the battery cover and then possibly penetrate the protective casing of the lithium-ion battery which may result in a fire. there have been no reported incidents of fire, injury or death due to this potential safety hazard. a correction has been developed to eliminate the potential risk. thecorrection alters the design of the battery cover, creating a barrier that prevents mounting screws from penetrating into the battery compartment.
  • 조치
    A Mortara Instrument Australia service representative will be contacting affected customers to schedule an update that includes replacement of the battery cover and replacement of the battery pack if mounting screw damage is evident. This update is at no cost to the customer. This action has been closed-out on 1/03/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Surveyor S12 and Surveyor S19 Patient MonitorsMultiple part numbers and serial numbers affectedManufactured March, 2012 to September, 2014ARTG Number: 215421
  • Manufacturer

Manufacturer