Symbia E and e.cam collimator (nuclear imaging gamma camera system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00330-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens have received complaints regarding incidents where collimator change activities have resulted in operator injury. the investigation found that these incidents may have been avoided by operators properly following specified instructions. the system was designed to safely store and exchange collimators, however if the instructions are not followed serious injuries, such as broken bones and lacerations, may occur. siemens have determined that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasise the proper collimator change process.
  • 조치
    Siemens is providing users with addendum instructions to follow during collimator change activities. This action has been closed-out on 27/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Symbia E and e.cam collimator (nuclear imaging gamma camera system)Catalogue Numbers: 10275879, 10413009, 04380213, 04380221, 05242826, 05977066, 05977074, 05984005, 05989079, 05989087, 05989095, 05991109, 05991117, 05992099, 07324119, 07324135, 07324143, 07324150, 07760809, 07760932, 07761161, 07823920, 07823938, 07823946, 07823953, 07823961, 07823979, 10151531, 10151532, 10520745ARTG Number: 186317
  • Manufacturer

Manufacturer