Symmetry Surgical/Olsen – Single Use Electrocautery Forceps/Pencils/Cords 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00795-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Device technologies has been notified by the manufacturer, symmetry surgical, of a product complaint regarding a monopolar pencil and cord. the reporting site observed some anomalies in the packaging seal resulting from the package’s contents having migrated near or within the seal area. this condition increases the risk of the seal’s integrity being compromised. testing indicated that the defect is introduced during the packaging operation and will not result from shipping or normal handling conditions.
  • 조치
    Device Technologies (DTA) is advising customers product is acceptable for use if it is visually confirmed that the package's contents do not interfere with the packaging seal and the seal is intact. Users are advised to either: - Inspect all affected product in accordance with the inspection instructions outlined in the Customer Letter. Product that does not meet acceptance criteria should be shipped back to DTA for credit. If users are uncertain if product is acceptable it is advised to return the item; OR - Ship stock held of the Olsen Single Use Electrocautery Forceps/Pencils/Cords back to DTA for credit.

Device

  • 모델명 / 제조번호(시리얼번호)
    Symmetry Surgical/Olsen – Single Use Electrocautery Forceps/Pencils/Cords ARTG Numbers: 214571, 214834, 214569
  • Manufacturer

Manufacturer