SynchroMed II and SynchroMed EL Implanted Infusion Pump 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00636-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-07-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Within the synchromed pump, feedthroughs are components that provide an electrically insulated path for current to flow from the electronic circuitry to the motor. an electrical short can occur when ions from the drug solution and humidity permeate through the drug pathway tubing inside the pump and interact with the feedthrough over time. an electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and leads to a loss of or reduction in therapy which may result in the return of underlying symptoms and/or withdrawal symptoms.
  • 조치
    Through the 'Hazard Alert' letter, Medtronic is advising clinicians regarding ongoing patient management recommendations. Medtronic does not recommend prophylactic replacement of SynchroMed II or SynchroMed EL pumps due to the estimated low occurrence rate, the presence of pump alarms, and the risks associated with replacement surgery. However, appropriate consideration should be given to individual patient needs. If repeated short duration motor stalls, Low Battery Resets (with or without Safe State), or a premature Elective Replacement Indicator occur, replacement surgery should be scheduled for therapy continuation. Alternative medical management should be considered if appropriate. For more details, please see http://www.tga.gov.au/safety/alerts-device-synchromed-130718.htm .

Device

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.