SynchroMed II and SynchroMed EL Implanted Infusion Pump 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00678-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-07-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The clinician refill reference card for synchromed implantable infusion systems that was originally distributed with the january 2011 safety alert (tga ref.: rc-2010-rn-01266-3) related to pocket fills has been updated to align with new product labelling. the january 2011 medical device correction letter provided important reminders concerning the potential for a pocket fill during a synchromed ii or synchromed el implantable drug pump refill procedure, and important patient management recommendations. a pocket fill is the inadvertent injection of all or some of the prescribed drug into the patient’s subcutaneous tissue, which includes the pump pocket, instead of the pump which can lead to life-threatening symptoms, serious patient injury, or death due to overdose or underdose.
  • 조치
    The main title of the Clinician Refill Reference Card has been updated to read Critical Actions in the Pump Refill Procedure, and the updates to the card include: - A description of the card’s purpose regarding pocket fill - A reminder to clinicians of the critical steps for ensuring the pump is correctly refilled - Detail regarding proper alignment of the refill template - Information for actions to take if a pocket fill is suspected - Removal of the note related to glucose testing Medtronic has updated product manuals and is in the process of deploying the updated labelling. Current labelling for product manuals can be found at www.medtronic.com/manuals. For more details, please see http://www.tga.gov.au/safety/alerts-device-synchromed-130718.htm .

Device

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.