Synchron Systems Enzymatic Creatinine (CR-E) and Uric Acid (URIC) Reagents. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01017-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2015-10-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Beckman coulter has confirmed that n-acetyl cysteine (nac) used in the treatment of acetaminophen overdose may cause assay interference when administered in therapeutic concentrations with assays using the trinder chemical pathways; enzymatic creatinine (cr-e) and uric acid (uric). nac interference may lead to falsely low results for uric acid and creatinine in patients who are concurrently being treated for acetaminophen overdose with nac.
  • 조치
    Beckman Coulter is amending the INTERFERENCES section of the Chemistry Information Sheets to include N-Actel Cysteine. There is no requirement to review previously generated results. This action has been closed-out on 04/03/2016.