syngo Dynamics v9.5 with Sensis-syngo Dynamics Integration (Radiology picture archiving and communication system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00425-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-04-17
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Sensis reference numbers greater than 1,000,000 are wrongly modified during transfer to syngo dynamics. in this case during “in-progress” a study will get a sensisrefno for the wrong study in sensis. this will cause an incorrect link between syngo dynamics images and the sensis study data. when a patient study is opened in syngo dynamics a sensis report of another patient study is shown.The preconditions for this to occur are:- sensis-syngo dynamics integration- sensis study is opened from syngo dynamics while still "in progress"- sensis reference number is > 1000,000this could result in patient mismatch or a persistent unintended change to patient data could result. the patient mismatch is detectable when the study is opened for the first time if the toolbar is not configured to auto-hide.
  • 조치
    Siemens is preparing a service patch modification that will resolve the potential malfunction. In the interim, end users are requested to discontinue opening "in progress" studies and to configure the “AXIOM Artis AE” title in syngo Dynamics to "Other" type of station. This action has been closed-out on 02/02/2016.