Synthesis, Avant & Primo2x Families Adult Oxygenator Mounting Brackets (used with Sorin 3T Heater-Cooler devices) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 LivaNova Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00886-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Livanova (sorin) have found that certain oxygenator mounting brackets that include tubing that connects the heater-cooler 3t water circuit to the oxygenator could be a source of re-contamination (with non-tuberculous mycobacteria) of a clean heater-cooler 3t’s water circuit. this tubing is not subjected to disinfection as per the ifu for the heater-cooler. if the heater-cooler 3t user failed to clean and disinfect the heater-cooler 3t resulting in contamination of the heater-cooler 3t and at some point later used oxygenator mounting bracket, the water from the contaminated heater-cooler 3t could cross-contaminate the oxygenator mounting bracket’s internal tubing in which biofilm could develop. if a non-contaminated heater-cooler 3t was later connected to an oxygenator mounting bracket that had been previously cross contaminated, the oxygenator mounting bracket could contaminate the clean heater-cooler 3t.
  • 조치
    LivaNova will replace the affected bracket and is providing advice to users on how to dispose of the affected Oxygenator bracket. The existing brackets can be used until a replacement unit can be provided.

Device

  • 모델명 / 제조번호(시리얼번호)
    Synthesis, Avant & Primo2x Families Adult Oxygenator Mounting Brackets (used with Sorin 3T Heater-Cooler devices)D632 Adult Open System Bracket Catalogue Number: 05046Adult Oxy module Avant Bracket Catalogue Number: 050128Synthesis “C” BracketCatalogue Number: 050350Synergy Bracket Catalogue Number: 050131D626 Avant Holder Catalogue Number: 05040D630 Avant Holder Catalogue Number: 05041Synthesis Oxygenator Holder Catalogue Number: 050127ARTG # 194514
  • Manufacturer

Manufacturer