Sysmex XN and XN-L Series - All Sysmex XN analysers versions 16 to 19 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01165-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2015-12-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Roche diagnostics has been informed by sysmex that incorrect data transfer to the host computer may sporadically occur, when using the sysmex xn analysers, versions 16 to 19. this may lead to delays in reporting results and false negative results. under certain situations data may not be sent to the host computer, or incorrect data may be transferred. an example of incorrect data transfer may be if data is transferred as wbc, rbc, aso%, neut# and if this occurs at the transfer of baso% data, then all of the subsequent item’s data will be overwritten as baso% data. it has not yet been established which item overwrites the subsequent item’s data. if multiple samples were selected, and those selected samples were sent to the host computer as batch data, then this issue may be observed on multiple samples.
  • 조치
    Roche Diagnostics will perform a software update on the affected devices. Roche is advising the users to investigate result transmission for data integrity if the error messages “TCP/IP send error occurred at host computer” or “Unable to establish TCP/IP connection with host computer” have been noticed prior to the software patch being installed. This action has been closed-out on 05/09/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Sysmex XN and XN-L Series - All Sysmex XN analysers versions 16 to 19ARTG Number: 173886An in vitro diagnostic device