Thermo Scientific Remel Neisseria meningitidis Poly A-D 2mL and Neisseria meningitidis Group D 2mL. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Thermo Fisher Scientific Australia & New Zealand 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00201-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-03-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An internal technical investigation by remel europe, part of thermo fisher scientific, has confirmed that remel neisseria meningitidis poly a-d (r30166601) and neisseria meningitidis group d (r30167001) may fail to agglutinate within the specified minimum reaction time when tested with group d neisseria meningitidis bacteria. continued use of these lots may result in a failure to correctly identify isolates as group d neisseria meningitidis bacteria. the product thermo scientific remel neisseria meningitidis poly a-d (r30166601) continues to correctly identify neisseria meningitidis groups a, b and c. no other batches or pack sizes of this product are affected by this recall.
  • 조치
    ThermoFisher is advising users to inspect stock and remove all units of the affected batches from further use. Users should complete the Facsimile Reply Form provided with the Customer Letter to arrange for a credit note to be issued.

Device

  • 모델명 / 제조번호(시리얼번호)
    Thermo Scientific Remel Neisseria meningitidis Poly A-D 2mL and Neisseria meningitidis Group D 2mL. An in vitro diagnostic medical device (IVD)Catalogue Numbers: R30166601, R30167001Lot Numbers: 2239498, 2177820, 2239861, 1773410 & 1739963. ARTG Number: 236824(Thermo Fisher Scientific Australia & New Zealand - Microbial-isolate agglutination reagent IVDs)