ThinPrep 5000 Processor with Autoloader 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hologic Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00516-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An investigation of customer complaints has determined that in rare cases, it is possible for the thinprep 5000 processor with autoloader to misread the value encoded in the vial barcode label. this occurs when the barcode scanner decodes a value that meets all symbology rules and check-digit validation but is not the intended value encoded in the barcode and processes the vial using this identifier. this potential is only present in thinprep 5000 processor with autoloader configurations and is not present in the benchtop configuration. in an autoloader configuration, if a misread occurs, the misread vial id value is transferred to the slide and a slide is produced with the incorrect identifier. there is a risk of misidentification of a patient slide.
  • 조치
    Hologic is advising users to ensure the labels are graded against the appropriate standard. The Technical Information Reference Guide has been updates to clarify the barcode quality and specification requirements. Hologic is providing users with a software update to perform additional safety reads to improve identification of inadequate labels. Hologic is also providing users with updated barcode scanners. This action has been closed out on 19/01/2016.