TOPCON – 3D Optical Coherence Tomography 3D OCT-1 MAESTRO machines 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Device Technologies Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00517-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-06-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In certain usage environments, when the 3d oct-1 is used over a prolonged period of time, dust can penetrate into the optical system and adhere to the surface of the lens. this dust accumulates gradually over time and this can cause a whitish shadow artefact to occur in centre of the image and potentially create a blurry background. if this occurs, the reduced image quality is an obvious anomaly noticed during clinical assessment which cannot be mistaken for patient pathology.
  • 조치
    Topcon Corporation will perform a retrofit to block the penetration points to prevent dust from entering into the optical system. In order to do this, Device Technologies will organise the return of affected machines to the manufacturer for the upgrade. A loan unit will be provided for the site whilst their machine is being upgraded. The corrective retro-fit action will be performed in a systematic manner, over a period of time due to the limited number of loan units available. Although affected systems may not exhibit any evidence of this artefact, it is recommended that all units undergo this preventative measure. Customers may continue to use their machine as usual until a loan unit is available for the site.

Device

  • 모델명 / 제조번호(시리얼번호)
    TOPCON – 3D Optical Coherence Tomography 3D OCT-1 MAESTRO machines Product code: TOP-470109800ARTG Number: 135436
  • Manufacturer

Manufacturer