TRUEresult Blood Glucose Test Strip (10 counts) (used with the TRUEresult Blood Glucose Monitoring System). An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Nipro Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00632-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-05-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This recall has been initiated as a limited number of sealed test strip vial boxes from the affected trueresult blood glucose test strip batch may include an open test strip vial. test strips contained within open vials are exposed to the outside environment, which can adversely affect the strips performance. in particular, test strips exposed to the environment may provide falsely low blood glucose results. falsely low blood glucose results may lead to undetected hyperglycemia (high blood glucose) and result in improper treatment.Action based on falsely low results may raise blood glucose levels further, resulting in potentiated hyperglycemia which may require medical intervention.
  • 조치
    Nipro is advising users to inspect stock and quarantine any remaining stocks of the affected batch for return. This action has been closed-out on 03/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    TRUEresult Blood Glucose Test Strip (10 counts) (used with the TRUEresult Blood Glucose Monitoring System). An in vitro diagnostic medical device (IVD) Item Number: E4i08-81FDExpiry Date: 25/06/2018Batch Number: KS1741ARTG Number: 180121
  • Manufacturer

Manufacturer