TruVidia Wireless Receiver 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Trumpf Med Aust Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00075-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-01-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Trumpf medical discovered that the radiated emission from the truvidia wireless receiver is above the normative accepted limits. it is unlikely, but possible, that the radiated emission may interact with other devices in the operating room, which may cause other devices to not function properly. the frequency of interference and associated risk will be dependent upon the immunity of the other devices for radiated emissions and nature of the failure mode of the operating room equipment being used at the time.
  • 조치
    Trumpf Medical is advising users that an upgrade of the affected components will be completed as a permanent correction. If users suspect that the device may be interacting with other devices in the operating room, switch off the wireless receiver (ensure the receiver is switched off, it is not sufficient to switch off the camera or control unit).

Device

  • 모델명 / 제조번호(시리얼번호)
    TruVidia Wireless Receiver Model Number: 1940747Serial Number: 102584474 to 103214022ARTG Number 274674
  • Manufacturer

Manufacturer