Tumor Localization (TumorLOC) software application used on Philips PET/CT combination diagnostic imaging systems PET/CT system with TumorLOC supplied in Australia: GEMINI TF 64 Slice 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00627-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-07-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has identified these issue with the tumorloc software1) when a single contour is saved in edit mode, tumorloc does not save the contour in the correct location. the contour is saved in the location of the same slice where the last contour was drawn for the last region of interest (roi) in the roi list. a misplaced contour then results in a misplaced isocenter or incorrect field size. 2) when the region of interest (roi) is displayed on the beam’s eye viewer (bev) in outline mode, and the roi colour is changed, the colour is changed in the multi-planar projection (mpr) viewers but not the bev viewers. 3) tumor loc-created datasets are flipped and incorrectly labelled when non-gated data is loaded along with pulmonary gated data and is used to create intensity projection datasets (mip, minip, or avgip).4) duplicate roi or isocenter names can be created in tumor loc. this duplication is only possible when using the copy feature or manually naming a roi with the suffix copy.
  • 조치
    A Philips Field Service Engineer will implement the Tumor LOC software upgrade to the affected system. In the interim, users are advised to implement workarounds to prevent the failure mode from occurring. This action has been closed-out on 15/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Tumor Localization (TumorLOC) software application used on Philips PET/CT combination diagnostic imaging systems PET/CT system with TumorLOC supplied in Australia: GEMINI TF 64 SliceARTG Number: 118077
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA