Ultrasonic Transducer Kits used with the Access Family of Immunoassay Systems. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00562-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-05-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Beckman coulter has determined that some ultrasonic transducer kits were potentially assembled with incorrect mounting screws. one of these transducers may have been installed on the system during a recent instrument repair. over time the screws can become loose if a combination of correct and incorrect screws are used and possibly cause incorrect probe wash, splashing in the reaction vessels (rv), incorrect particle suspension in the reagent pack, or incorrect rv mix due to alignment of the probe to wash tower/rv leading to assay imprecision. this may cause erraneous but believable results and/or delay of results due to transducer alignment error.
  • 조치
    Users are advised to review historical QC data and System Check results since the transducer was installed for any indication of a potential problem and determine whether a review of patient test results is warranted. A service representative will be contacting users to schedule a service visit to confirm that their transducer is properly installed.

Device

  • 모델명 / 제조번호(시리얼번호)
    Ultrasonic Transducer Kits used with the Access Family of Immunoassay Systems. An in vitro diagnostic medical device (IVD)
  • Manufacturer

Manufacturer