Unicel DxH Slidemaker Stainer Coulter Cellular Analysis System. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00808-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-06-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Beckman coulter has determined that a fitting installed in a select population of slidemaker stainer (sms) units may not completely seal after being installed. please note that not all fittings installed may be affected. for those affected, one may observe a slow dripping leak of the staining reagents (methanol, alcohol and buffers) into the spill tray below designed for this purpose. there is no impact to patient results due to this issue. there is no new or additional safety risk associated with this issue beyond the continued potential of exposure to stain and methanol associated with the handling of these chemicals resulting from a slow dripping leak.
  • 조치
    Beckman Coulter is asking that customers: 1 - Immediately contact your Beckman Coulter representative to schedule a service visit in the event of a leak. Beckman Coulter Field Service Engineers will replace the fitting on the impacted instruments. 2 - Inspect the reagent supply drawer for spills; if necessary, clean up with absorbent materials such as paper towels, and dispose in accordance with your local regulations and laboratory procedures. 3 - Follow best laboratory practices when using the instrument, which includes the use of appropriate personal protective equipment (PPE) to protect the user from exposure.

Device

  • 모델명 / 제조번호(시리얼번호)
    Unicel DxH Slidemaker Stainer Coulter Cellular Analysis System. An in vitro diagnostic medical device (IVD)Software version 3.2Part Number: PN 775222ARTG Number: 177999
  • Manufacturer

Manufacturer