USB Communication Data Loss – Handheld Base Unit and Accu-Chek Inform II Base Unit 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Roche Diagnostics Australia Pty Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00221-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-03-01
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Some handheld base units (hbu), used for coaguchek professional meters, cobas h 232 meters and usb communication data loss – handheld base unit, might produce transmission errors in the form of data loss in the usb communication between the meter and the dms. this could happen in rare cases when the base unit is connected via usb connection to a pc. this may be indicated by numerous (nonspecific) error messages sent from the meter to the dms in the form of “unexpected sw failure 11800 events.This issue can lead to the data loss, or in the worst case, to an erroneous assignment of the patient data (patient mismatch). in case of a patient mismatch, taking into account difficult detectability of this issue, a serious medical adverse event cannot be excluded.
  • 조치
    A mandatory software update for the “Roche HBU/BU Control Panel” will be required once released by Roche. Until this is available, please do not use the USB connectivity option without further precautions. Roche suggests implementing one of the below workarounds: ·Switch the connectivity method to Ethernet (TCP/IP) or Wi-Fi. ·Install a USB hub or an active USB-repeater to improve the USB signal, to significantly reduce the potential transmission errors. One of these is required per USB connection and the software update will still be required. ·Due to some variances in Base Unit and computer hardware, only specified USB hubs or active USB-repeaters can address the issue - please ask Roche Diagnostics Australia for information. This action has been closed-out on 30/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    USB Communication Data Loss – Handheld Base Unit and Accu-Chek Inform II Base UnitVarious affected productsARTG number: 202785 (cobas h 232 POC system)ARTG number:197926 (CoaguChek Professional meters)
  • Manufacturer

Manufacturer