V5Ms transesophageal transducer (used for ultrasound imaging) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00943-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-08-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Seiemens has received recent overseas reports about the deterioration of material covering the articulating section of the v5ms transesophogeal transducer. although to date, no patient injury has been reported, a compromise or fault in this material combined with a failure to use a probe cover as directed in the user manual may cause oesophageal cuts, bleeding, and perforation to the patient, and severe trauma, electrical burns, and serious electrical hazards to the patient and the user. however, a leakage current test can indicate cracks, cuts, tears, perforations, and protrusions in the articulation material.
  • 조치
    To avoid the potential risk associated with this issue, customers are asked to undertake the following steps: - Perform the leakage current test for a transesophageal transducer prior to each use. - Use only Siemens approved cleaning and disinfecting solutions. - Use market-cleared transducer sheaths specifically designed for TEE applications. - When users store the transesophageal transducer, take extreme care that the distal tip is straight. Never move the distal tip by hand. Always use the flex controls. Customers are advised that if they notice any damage or wear on the transducer, or if the transducer fails to pass the leakage current test, immediately discontinue use of the transducer and contact their local Customer Service Engineer. This action has been closed-out on 06/06/2016.