Various Microgenics Application Sheets. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Thermo Fisher Scientific Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00692-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-06-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During an internal assessment, thermo fisher scientific found that application sheets provided by the manufacturer microgenics for expanded use of some assays are inadequate. specifically, the appropriate validation data for the expanded use of these assays, as indicated in the instructions for use (ifu) when not used on manufacturer approved analysers, is inadequate. in addition, microgenics did not clearly state in these application sheets that the parameters provided are only to be used as starting points and that each assay setting needs to be validated individually by the user.Thus the information given on the specific application sheets for use with non-microgenics approved analysers may be ineffective. it is possible that assays not validated for use with other instrument platforms could potentially lead to erroneously low or high test results.
  • 조치
    Thermo Fisher Scientific is requesting users to 1) immediately discontinue use of the affected application sheets used on their specific instrument platform, 2) review their internal assay validations to determine if their current instrument settings using these assays require additional action within their clinical laboratory settings (consult a qualified healthcare professional, as appropriate) and 3) contact Customer Service to check if additional action is required.

Device

  • 모델명 / 제조번호(시리얼번호)
    Various Microgenics Application Sheets. An in vitro diagnostic medical device (IVD)Multiple Assays / Instrument PlatformsARTG Numbers: 192621 and 235822
  • Manufacturer

Manufacturer