  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01213-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-11-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An investigation has found that the units manufactured since april 2013 were not manufactured according to specification. this manufacturing issue may cause an electrical short in the generator. the issue could result in excessive rf energy being delivered to the renal artery with intermittent temperature spikes that may range between 86 - 93 degrees c.
  • 조치
    Customers are requested to segregate the devices immediately and organise for the return to Boston Scientific.

Device

  • 모델명 / 제조번호(시리얼번호)
    Vessix Generator System (portable RF generator used with percutaneous angioplasty balloon catheter to deliever low frequency RF energy into the renal artery)Material Number: H749RDNS0100Catalogue Number: RDNS010Serial Numbers: 42041, 42020, 42011, 42034, 42042, 42035ARTG Number: 208743
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”