VisuMax 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Carl Zeiss Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01144-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-09-04
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Based on post market surveillance activities, carl zeiss has received information that in certain devices there is an increased probability of failure of the bpc board. the function of this board is to control an electrically driven mirror which adjusts the laser beam position if necessary. this design allows a position compensation in case the laser beam should drift.It has been determined that there was a deviation in the soldering process during the manufacture of two batches of the bpc board which can affect the reliability of the visumax beam pointing control subsystem.The probability of injury is extremely unlikely. any failure of the bpc board like error 0x3172 is detected by the monitoring functions of the device either during startup or before start of a treatment. in case of a system error the system test is cancelled before any treatment. if an error occurs, the device goes in a safe mode.
  • 조치
    Complete and return the supplied acknowledgement form. Carl Zeiss service staff will contact users to arrange an appointment for the BPC Board exchange.

Device

  • 모델명 / 제조번호(시리얼번호)
    VisuMaxCatalogue Number: VISMAXSerial Number: 1173839ARTG Number: 150527
  • Manufacturer

Manufacturer