VITEK 2 60/XL and VITEK 2 Compact systems (In vitro diagnostic medical devices (IVD) used Bacterial identification and antibiotic sensitivity system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Biomerieux Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00168-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-02-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Under specific conditions, vitek 2 60/xl and vitek 2 compact pc systems connected to a laboratory information system have the potential to link a test result to an incorrect patient record. the four conditions required to potentially cause an occurrence of this issue are:- patient demographics option is enabled- patient demographics are not associated with the isolate- user does not resolve qualified isolates- laboratory re-uses lab ids (accession numbers) prior to qualified isolates being resolved.
  • 조치
    Biomerieux is updating the instructions for use (IFU), addendum to the software user manual, to include appropriate instructions for resolution of Qualified isolates as a short term action. Biomerieux is implementing a software solution to reduce the likelihood of occurrence of this issue in the next software update, version 7.01 targeted for launch in November 2013.

Device

  • 모델명 / 제조번호(시리얼번호)
    VITEK 2 60/XL and VITEK 2 Compact systems (In vitro diagnostic medical devices (IVD) used Bacterial identification and antibiotic sensitivity system) Used with the following PC models:Hewlett-Packard rp5700, Catalogue Number W0452Hewlett-Packard rp5800 XPE, Catalogue Number 413642Hewlett-Packard WES7, Catalogue Number 413862Hewlett-Packard dc7800, Catalogue Number W0449Hewlett-Packard dc7700 8-port, Catalogue Number W0447Hewlett-Packard dc7100 8-port, Catalogue Number W0441
  • 의료기기 분류등급
  • Manufacturer

Manufacturer