VITROS 3600 Immunodiagnostics System and VITROS 5600 Integrated System, software version 3.2.2 and below. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Ortho-Clinical Diagnostics 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00420-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-04-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A vitros system software timing anomaly has been identified that could cause two different sample metering scenarios that may lead to erroneous results.Scenario 1: the vitros system could aspirate sample from an unintended sample container causing assay result(s) obtained from that sample to be incorrectly associated with the intended sample.Scenario 2: a sample could be aspirated from a sample container (sample a) and be dispensed into an unintended container (sample b) to be contaminated and diluted by sample a.There is the potential for false negative and false positive results which may lead to an inappropriate diagnosis and patient management leading to serious patient injury.
  • 조치
    Ortho is providing users with interim instructions to follow to decrease the probability of the issues occurring, and can review the affected systems to determine if the errors have occurred. It is recommended to discuss any concerns regarding previously reported results with the Laboratory Medical Director to determine the appropriate course of action. Ortho will be providing users with a software update (software version 3.2.3) as a permanent correction. This can be automatically downloaded for systems connected to e-Connect or software kits provided for systems not connected to e-Connect. This action has been closed-out on 20/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    VITROS 3600 Immunodiagnostics System and VITROS 5600 Integrated System, software version 3.2.2 and below. An in vitro diagnostic medical device (IVD)VITROS 3600 Immunodiagnostics System, software version 3.2.2 and belowProduct Code: 6802783Unique Device ID: 10758750002979VITROS 5600 Integrated System, software version 3.2.2 and belowProduct Code: 6802413Unique Device ID: 10758750002740ARTG Number: 180199