VITROS Chemistry Products UPRO Slides for measurement of protein in urine (using VITROS 250/350/5,1 FS/4600 and 5600 Systems). An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Ortho-Clinical Diagnostics Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01627-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-14
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ortho clinical diagnostics (ortho) has identified the potential for biased urine protein results using vitros upro slides. the following observations were made when using vitros upro slides:-1. variability gen to gen for patient samples. impact to results: magnitude of bias is dependent upon the gen;-2. variability in results for random and 24-hour urine samples with low specific gravity. impact to results: unexpected increase in the number of results exceeding the reference intervals for random and 24-hour urine samples.; and-3. negatively biased results. impact to results: negatively biased results versus the ortho pyrogallol red comparative method.Ortho is providing users with the results of their internal testing that details the levels of bias observed for each of the scenarios above.Gens 12-18 are affected, however the gen 12 and 13 product have expired.
  • 조치
    Ortho is advising users to immediately discontinue using and discard all VITROS UPRO Slides remaining in their inventory. Ortho will credit accounts for any discarded VITROS UPRO Slides and VITROS Chemistry Products Calibrator Kit 10 and/or VITROS Chemistry Products UPRO Performance Verifier I & II. Any concerns users may have regarding previously reported urine protein results should be discussed with the Laboratory Medical Director to determine the appropriate course of action. As Ortho is unable to provide replacement VITROS UPRO Slides at present, they recommend laboratories use an alternative method until further notice by Ortho of future product availability.

Device

  • 모델명 / 제조번호(시리얼번호)
    VITROS Chemistry Products UPRO Slides for measurement of protein in urine (using VITROS 250/350/5,1 FS/4600 and 5600 Systems). An in vitro diagnostic medical device (IVD).VITROS UPRO Slides GENs 14, 15, 16, 17 & 18Product Code: 6800120Expiry Date range: 1 Jan 2017 to 1 May 2018ARTG Number: 232251
  • Manufacturer

Manufacturer