VOLISTA STANOP, TRIPOD, ACCESS and QUICKLOCK Light heads 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getinge Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00656-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-05-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Getinge has identified that a screw is missing from the racket of the cupola on some devices in production. it is not yet confirmed whether the screw was missed during manufacture or had loosened over the life of the device. as a consequence the light head shaft could break and the cupola could detach itself when manipulated and be held only by its cables. it appears such event is more likely to happen mainly during maintenance, cleaning and surgery preparation.
  • 조치
    Getinge are advising they will contact affected sites to plan for update of devices. In the interim, users should verify whether the cupola has mechanical play by following the guidance provided in the Customer Letter. If no play in the cupola is detected the device may continue to be used. If play is detected, users should cease use of the device and contact Getinge immediately.

Device

  • 모델명 / 제조번호(시리얼번호)
    VOLISTA STANOP, TRIPOD, ACCESS and QUICKLOCK Light heads All units manufactured until December 2016Light Head Part Numbers: ARD568811950, ARD568811960, ARD568811961, ARD568811962, ARD568812910, ARD568812950, ARD568812960, ARD568812962, ARD568821913, ARD568821960 and ARD568822960ARTG Number:162037