VueOptic Visualisation Source & Flexor Vue Deflecting Endoscopic System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 William A Cook Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00695-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-05-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection and sterilisation of these products. in addition, the monitoring of chemical residues was not considered in the reprocessing validation. preliminary investigation indicates that validation data related to the reprocessing of these devices do not meet the current guidelines. potential adverse events that may occur if the products are not adequately reprocessed include urological infections and systemic infections from a urological origin as well as events resulting from chemical residual exposure. to date, there have been no reports of adverse reactions related to inadequate cleaning, disinfection, sterilisation or chemical residual exposure associated with these devices.(please note that for the endoscopic system, the reprocessing instructions apply to the vueoptic visualisation source as the flexor 180 is intended for one-time use only).
  • 조치
    Cook Medical will be updating the reprocessing instructions in the Instructions For Use (IFU). In the interim, users are advised to quarantine any affected products. Use of the affected products can be continued if the updated "Reprocessing Instructions" provided is adhered to. (Users are to note that the reprocessing instructions apply to the VueOptic Visualisation Source as the Flexor 180 is intended for one-time use only).

Device

  • 모델명 / 제조번호(시리얼번호)
    VueOptic Visualisation Source & Flexor Vue Deflecting Endoscopic SystemVueOptic Visualisation Source Catalogue Identifier: FVO-150Part Number: G25343Flexor Vue Deflecting Endoscopic System Catalogue Identifiers: FV-090075-150 & FV-090045-150Part Numbers: G50972 & G34306 (respectively)All Lot NumbersARTG Numbers: 211575 and 211572
  • Manufacturer

Manufacturer