Zimmer A.T.S. 3000 Automatic Tourniquet System (used to temporarily occlude blood flow during surgery) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00980-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-09-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    On some occasions during surgical procedures, the unit will fail, alarm, lock into a non-operational mode and the display screen will go blank. based on zimmer's investigations, this problem is due to a defective chip on the control board.
  • 조치
    Customers are being asked to remove affected A.T.S 3000 tourniquet systems from use immediately & contact their sales representative who will return the unit to the company for repair and arrange a loan unit.

Device

  • 모델명 / 제조번호(시리얼번호)
    Zimmer A.T.S. 3000 Automatic Tourniquet System (used to temporarily occlude blood flow during surgery)Product Code: 3011LACWSerial Numbers: 3011LACS, 3011LACL, 3011LAED, 3012BAAB, 3012BAAF
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA