Zimmer Air Dermatome II Handpiece (used during skin grafting surgeries) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00976-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Zimmer biomet surgical has received product complaints reporting that the surface coating applied to the device is blistering, peeling and discolouring after usage over time. there have been no injuries associated with the reported complaints. in addition, an independent assessment determined that the blistering, peeling and potentially flaking of the surface coating is not expected to result in any long range health consequences. given that the frequency of this surface coating condition can worsen with age, zimmer biomet is currently making preparations for replacement activities to follow in order to prevent future complaints.
  • 조치
    Zimmer is advising users to inspect the devices before and after each use. If blistering or peeling is observed, discontinue use of the affected device. If an alternative device is not available, at the surgeon's discretion, determine the use of the device as a medical necessity and continue to use. After use, follow the facilities wound irrigation protocol or Zimmer Biomet recommends copious irrigation with a normal saline solution or immersion of the graft in a bowl of normal saline. Zimmer Biomet recommends irrigating the donor site with saline as well. Carefully examine the skin graft and donor site for grey coating flakes and continue to irrigate as necessary. All affected devices will be replaced with unaffected stock when available. This action has been closed out on 20/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Zimmer Air Dermatome II Handpiece (used during skin grafting surgeries)Item Number: 00-8851-001-00All serial numbers affectedARTG Number: 115929
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA