Zimmer Persona Tibial Articular Surface Provisional (TASP) Instruments (Supplied as part of the reusable surgical kit used to aid the surgeon in assessing the thickness of articular surface required) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01033-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-09-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Zimmer has received complaints related to the fracture of persona tibial articular surface provisional (tasp) tops and bottoms. based on the investigation results of these complaints, zimmer determined that updates to device labelling associated with the zimmer persona knee system were required to provide additional clarifications relating to the instructions for use of the tasp construct. .
  • 조치
    Zimmer is advisingthat the updated technique for use of the TASP construct and its components can be found on pp. 34-37 of the Persona Surgical Technique (97-5026-001-00, Rev. 8) which includes: • Apply gentle manual pressure without impacting the TASP construct with either a mallet or hand. The TASP construct includes the TASP top, bottom, shim, and Tibial Sizing Plate Handle. • During assembly of the TASP construct, slide the shim in using a direct anterior approach between the TASP top and bottom. To avoid inadvertent separation, maintain slight pressure between the TASP top and bottom while inserting the shim. • Varus/Valgus forces may make it difficult to remove the TASP construct. To aid in the removal of the TASP and prevent breakage, ensure that the joint is in a neutral position when removing the TASP construct. This action has been closed-out on 18/03/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Zimmer Persona Tibial Articular Surface Provisional (TASP) Instruments (Supplied as part of the reusable surgical kit used to aid the surgeon in assessing the thickness of articular surface required)Multiple part (item) numbersARTG number: 109936
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA