Zimmer Trabecular Metal Reverse Glenosphere Distractor(Supplied as part of the reusable instrumentation kit for shoulder replacement surgery) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zimmer Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00902-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-08-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Zimmer is initiating a recall of the trabecular metal reverse glenosphere distractor due to the potential of the device not functioning properly should it be utilised without proper lubrication. as a result, there is a potential for the device to fail to provide impact to the glensophere when the trigger is pulled. in addition, there is a potential for the foot of the device to fracture during use which may result in the device being unable to remove the glenosphere.
  • 조치
    Zimmer is requesting users to remove affected lots from use. Zimmer are replacing affected stock with replacement product which incorporates a new design.

Device

  • 모델명 / 제조번호(시리얼번호)
    Zimmer Trabecular Metal Reverse Glenosphere Distractor(Supplied as part of the reusable instrumentation kit for shoulder replacement surgery)Item number: 00-4309-049-00Multiple lot numbers affectedARTG Number: 109936
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA