ADVIA CENTAUR, Registration no. 10345160020 - Serial numbers - Attachment -Carta to the Client. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Siemens Healthcare Diagnósticos Ltda.; Siemens Healthcare Diagnostics Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1430
  • 날짜
    2014-09-26
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    The company informs that patient demographics (name, age, gender, ID) of an earlier sample are incorrectly transmitted to the LIS with the current patient sample results, the data also appears incorrectly in the worklist and printed reports . Patient outcomes are not affected if the ADVIA Centaur and ADVIA Centaur XP immunoassay systems use Sample ID (SID) as the primary identifier for all outcomes. A delay in reporting may occur. The overseas manufacturer performed the problem analysis, evaluating all the inherent risks that generated the corrective action in the field. Based on the investigation and mitigation of the risk conducted, this action was classified as without health risk, its severity being low and its probability of occurrence extremely unlikely.
  • 원인
    Siemens healthcare diagnostics has identified a problem with patient demographics sent to the lis of the advia centaur® / advia centaur® xp immunoassay systems. siemens has confirmed that under extremely rare circumstances the demographics of patients from a previous order received from the lis are merged with the next order. this problem may occur when the lis data buffer becomes full in the advia centaur system, and a particular character is encountered in the last five positions in the lis data buffer.
  • 조치
    Siemens recommends following the instructions provided in the chart if the equipment in use is connected to a LIS system that transmits demographic patient data at each request. If the equipment is connected to a LIS system that does not transmit demographic patient data, no action is required. In addition, Siemens recommends retaining the letter with the laboratory records and forwarding the notification to anyone who may have access to the products.

Manufacturer