Angiography equipment, Risk class: III, models described below. Model: ALLURA XPER FD10 (floor), ALLURA XPER FD10 (ceiling), ALLURA XPER FD20 (floor), ALLURA XPER FD20 (ceiling), Registration: 10216710153 ////. Model: INTEGRISH5000 F, Record: 10216710034 ////. Model: Integris Allura 12/15, Integris Allura 12 / 15F, Integris Allura 9C, Integris Allura 9F; Record: 10216710155 ////. Model: ALLURA XPER FD10 / 10; ALLURA XPER FD20 / 10; ALLURA XPER FD20 / 20; ALLURA XPER FD10 / 10 OR Table; ALLURA XPER FD20 / 10 OR Table; ALLURA XPER FD20 / 20 OR Table; Record: 10216710206 .. Serial number: 176; 553; 759; 1751; 001H021796; 001H021878; S4F010159A; S4F200174A; S4F200270A 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Philips Medical Systems Ltda.; Philips Medical Systems Nederland B.V. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1514
  • 날짜
    2015-02-05
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    According to the record holder's information, the risk is linked there is a possibility of collision of the equipment with the patient or with the staff present in the room. When suspended from the ceiling of the monitors, it may fall to its lowest position. There is also the possibility that the ceiling suspension of the monitors collide with other equipment that is close to it.
  • 원인
    According to the record holder, the ceiling suspending system of the equipment monitors may plummet to its lowest position which, after investigation pointed out that the cause of the failure was a montage suspended ceiling trigger assembly error.
  • 조치
    The registrant has established the following guidelines for customers until the containment action is implemented: - Do not position or move the ceiling suspension of the monitors on the patient - Do not allow team members to be under or near the suspension monitors.- Avoid any unnecessary movement of the monitors suspended ceiling. " (SEE ATTACHED)

Manufacturer