ARCHITECT CA 19-9 REAGENTS - Registration 80146501355 - Model ARCHITECT CA19-9XR - Lots Affected: 08849M500, 08851M500, 08852M500, 08853M500, 10122M500, 10040M500. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 ABBOTT LABORATÓRIOS DO BRASIL LTDA 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1147
  • 날짜
    2012-06-28
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    Abbott confirmed that six batches of the ARCHITECT CA19-9XR reagent show a change in the reported results for patients, and it was found that Abbott controls did not detect this change and presented results within the range. All other batches of reagents are not affected by this change. An internal investigation is under way. Initial indications suggest that the problem is caused by a component of the conjugate. All batches identified below have in common the same component of the conjugate. The investigation and collection of the impacted units in the client are in progress, and the forecast for completion is 09/15/2012. Anvisa follows this action.
  • 원인
    There may be a change in the reported results for patients, without abbott controls detecting this change.
  • 조치
    Customers who received batches impacted by this change were notified via an Abbott statement to suspend their use immediately. Abbott will replace the affected products.

Manufacturer