Commercial Name: Reactive Dry Vitros for ProteinC Reactive CRP / Dry Reactive Vitros for ProteinC Reactive CRP Technical Name: C Reactive Protein / Protein C Reagent ANVISA Registration Number: 10132590348 Hazard Class: II Affected Model: Box with 5 cartridges x 50 slides / Box with 5 cartridges x 18 slides Serial Numbers Affected: VITROS Chemistry Products PCR Slides; Product Codes: 1926740 and 8097990. Considering the scope of the affected lots (already distributed and possible commercialization), the list of customers using the inputs affected by this field action was considered based on the equipment purchased. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Johnson & Johnson do Brasil Ind. e Com. de Prod. para Saúde Ltda; Ortho-Clinical Diagnostics, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    2112
  • 날짜
    2016-11-28
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    Follow the instructions in the alert letter to perform dilutions when processing samples above the PCR measurement range. VITROS PCR Slides Instructions for Use (IFUs) will be revised to include additional instructions to follow when diluting samples with results above the measurement range. Meanwhile, keep the notification as verification of the reviewed process for processing samples above the measurement range.
  • 원인
    Ortho clinical diagnostics (ortho) received pcr reports with positive deviation when using vitros pcr slides for sample dilution. vitros chemistry products pcr slides products quantitatively measure c-reactive protein (crp) concentration in serum and plasma using the vitros 250/350 / 5.1 fs and 4600 systems and the vitros 5600 integrated system.
  • 조치
    Field Action Code 16000165 triggered under the responsibility of Johnson & Johnson do Brasil Ind. And Com. De Prod. for Health Ltda. Company will make correction in the field.