Commercial Name: Ureteric Illuminating Catheters Bush Technical Name: Catheters ANVISA Registration Number: 10330710065 Hazard Class: II Affected Model: 084120 and J-BICS-058020 Affected Lot / Serial Numbers: U2412468; U2412469; U2412470; U2413367; U2413911; U2429088; U2452670; 4,946,704; 4,965,697; 4,976,258; 4999997; 4,999,998; 5003247; 5007047; 4942636; 5,138,605; 5,141,718; 5,141,199; 5,145,105; 5,149,622; 5,149,623; 5,365,572; 5,365,642; 5,370,995; 5,388,926; 5509391; 5,518,548; 5,518,549; 5,593,065; 5,684,248; 5,718,485; 5,721,755; 5,737,801; 5,744,000; 5744001; 5,743,999; 5,766,046; 5,766,047; 5,789,746; 5794877; 5,798,213; 5,826,962; 5,973,018; 6008488; 5,978,226; 6008489; 6,056,601; 6073730; 6073731; 6127074; 6131751; 6131752; 6,142,020; 6,142,021; 6146513; 6146514; 6146518; 6228742; 6,242,198; 6371295; 6,375,171; 6,396,897; 6,396,898; 6214087; 6,247,244; 6255677; 6,262,720; 6,295,798; 6653171; 6719828; 6,519,829; 6,262,338; 6,360,647; 6,360,648; 6364334; 6371297; 6371298; 6,998,469; 7,012,893; 7,016,375; 7,016,376; 7,128,835; 7,134,480; 7162449; 7162450; 7162451; 7457006 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Handle Comércio de Equipamentos Médicos Ltda; Cook Incorporated 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    2245
  • 날짜
    2017-03-23
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    If direct skin or patient contact with overheated devices resulting in skin damage, including burns, it is recommended to seek medical attention.
  • 원인
    The cook medical group initiated voluntary recall of bush ureter illuminating catheters (references 084120 and j-bics-05802) due to the increased number of complaints that the black conductor in the proximal portion, which connects to the transparent component of the bush ureteral illumination catheter , it can overheat and melt. there have been no reports of overheating along the transparent component of the catheter, which is the part of the device that comes in contact with the patient's urethra, bladder, and ureter.
  • 조치
    Field Action Code No.207 / RNC 031 - 032 / RA 018 triggered under the responsibility of Handle Comércio de Equipamentos Médicos Ltda. Company will collect.