CRYSTAL ENTERICO NA FERMENTADOR (E / NF) ID KIT AND ACCESSORIES - Registration 10033430047 - AFFECTED LOTS: 2101435; 2104316; 2129487; 2163211. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 BECTON DICKINSON INDÚSTRIAS CIRÚRGICAS LTDA 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1223
  • 날짜
    2013-01-16
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    The product Enteric Crystal Non Fermenter ID Kit and Accessories is a miniaturized microorganism identification system, which employs conventional modified and chromogenic substrates. It is used for the purpose of identifying gram-negative aerobic bacteria belonging to the family Enterobacteriaceae as well as the most frequent isolates of gram-negative fermentative and non-fermenting glucose bacilli. Many of the tests used are modifications of the classical identification methods. This includes tests for fementation, oxidation, degradation and hydrolysis of various substrates. The research and development area of ​​the product factory evaluated the effect of the inverted position of Citrate and Malonic Acid substrates and found that the results presented were inconclusive either because they did not identify the microorganism or because they identified multiple microorganisms. In the latter case, the correct micro-organism always represented one of the results obtained. In cases of inconclusive results, the user will need to use another method or a combination of methods to be able to determine the correct identification of the microorganism. No incorrect identification results were found during the R & D tests or through reports of customer complaints. However, since the identification of other microorganisms that require the use of these two substrates can also be affected and because it is not possible to verify all possible results of the Crystal system, there is a remote chance of incorrect identification and, therefore, situation was classified as critical severity and of remote occurrence. The possible consequences of the use of the affected product are the failure to identify the microorganisms. Studies conducted by the R & D area; did not find incorrect results when the citrate and malonic acid substrates were inverted. Although it is not possible to state that an incorrect result will not occur based on the tests performed, it is more likely that in this situation an inconclusive result will be found for the identification of microorganisms. Anvisa follows this Field Action. UPDATE 08/23/2013 - The company finalized the collection of the remaining products in the market. Product has expired on 3/13/2013.
  • 원인
    Becton dickinson has identified that the product contains reagents citrate (cit) and malonic acid (mlo) in inverted positions. this may lead to inconclusive identification results, such as citrobacter and salmonella. however, the identification of other microorganisms that require the use of these two substrates can also be affected.
  • 조치
    The company recommends to professionals users that, verify the existence of the affected product in its stock; discontinue use immediately and promote disposal of remaining stocks. See Attached Customer Alert Message. http://en.wikipedia.org/w/eng/index.php