CULTURE BOTTLE BACT / ALERT MB - Record 10158120446 - Affected lots 1029125; 1029521. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 BIOMERIEUX BRASIL S.A 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1146
  • 날짜
    2012-06-25
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    BioMerieux detected that in the Instructions for Use for BACTAT bottles for BACT culture flasks, in Instructions for Use (document number 9300224A), the following steps were omitted: guidelines for the preparation of the MB Bact flask and the requirement to add 1 ml of enrichment to these bottles BioMerieux is providing to those who have received the affected product an Addendum to the Instructions for Use which includes the omitted information. A risk analysis has determined that the problem of "incorrect quantity or no addition of Enrichment Liquid" has a Moderate Risk, with Critical Severity and a Chance of Occurrence as Occasional. The probability was determined to be Occasional because the customer may make a pipetting error or omit the enrichment liquid during the preparation of the bottles. The probability of a lab error has not changed. Customers accustomed to the use of the product usually have protocols that the staff follows which must contain the appropriate volume of the enrichment liquid. Soon new customers would be more sensitive to this problem. However, new customers have the means to find the correct volume in the following locations: 1) in the Quality Control section of the product label, 2) on the bioMerieux website Bact Alert 3D product line - culture media table; 3) through the BioMerieux Customer Service. Based on the number of false negative claims since version 9300224A was released to the field, the absence of the amount of enrichment liquid volume does not appear to have negatively impacted customers and the risk remains Moderate. Consequently the risk in the field has not changed as a result of this problem. Anvisa is following this action.
  • 원인
    By omission of guidance due in the directions for use of the affected product, a failure to add the enrichment liquid may lead to a failure to culture the mycobacteria in specimens.
  • 조치
    Use the supplied Addendum and consult the Quality Control section in the Instructions for Use. You can also contact the BioMerieux Customer Service.