Equipment for angiography Artis. Model: Artis Zeego. Anvisa Registry n ° 10234230158. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Siemens Healthcare Diagnósticos Ltda. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1452
  • 날짜
    2014-10-24
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    The overseas manufacturer performed the problem analysis, evaluating all the inherent risks, which generated corrective action in the field. Based on the investigation and mitigation of the risk conducted, this action was classified as a medium risk to health, its severity being critical and its probability of occurrence extremely unlikely. According to the company, this is a possible hardware failure that has no influence on the treatment of patients. The customer will be advised to wait for the Service Team to schedule field corrections and to file the letter with the Operator's Manual. In addition, Siemens advises to inform all personnel of the organization who need to be aware of this problem and act accordingly. According to the company the field action was finalized
  • 원인
    A cable inside the c-arm of affected systems may have been improperly positioned. without an additional measure, there is the possibility of a cable inside the c-arm. the problem will not occur systematically but sporadically over the lifetime of the system.
  • 조치
    Correction in the field / Correction of parts or parts.