FABIUS MRI - ANESTHESIA WORK STATION. Anvisa Registry n ° 10407370065. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Drager Indústria e Comércio Ltda.; Drager Medical GMBH 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1685
  • 날짜
    2015-09-17
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    Dräger has been reported to have occurred in which the Fabius MRI or parts of the apparatus were attracted by magnetic resonance magnetic field. If the Fabius MRI is positioned too close to the resonance equipment, the magnetic force of the MRI can loosen parts of the Fabius MRI and injure people. The instruction manual and the marking of the apparatus specify that the equipment is to be used only in areas where the field strength does not exceed 40mTesla. However, the cases that were reported to the company resulted from the fact that the device was moved through the 40mTesla area and the marking of the device while it was not in use. For this reason, users are advised to make sure that the Fabius MRI is not moved or positioned in areas over 40mTesla (400Gauss), even at times when it is not used on the patient, in addition to following all information and warnings in the operating instructions of the equipment. In addition, users are advised to follow all the information that is presented during installation and training by Dräger's training team in Brazil.
  • 원인
    If the fabius mri is positioned too close to the resonance equipment, the magnetic force of the mri can loosen parts of the fabius mri and injure people.
  • 조치
    Send alert message to users and fix parts of the product.