HEMODIALISING MACHINE SPS-1550 (ANVISA Register: 10068390088). 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Baxter Hospitalar Ltda. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1018
  • 날짜
    2010-05-07
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    According to Baxter Hospitalar Ltda, as a result of a field correction performed in 2003 by the company, some hemodialysis equipment of the SPS 1550 model may contain electronic boards with wrong versions of software. The correct software versions for 1550 devices are 4.1B (for System 4 equipment) and 3.0A (for System 3 equipment). Software versions other than those listed above (on the side panel of the machine) display the incorrect "USE UF MANUAL" message when code errors FL04, FL07 and FL08 occur. Correct software versions (3.0A and 4.1B) display the message "DISCOUNT TX" (TX = Treatment) in the occurrence of these same error codes. The company has already sent a letter of communication and is making available the correct software update for its clients. #### Update: The action was closed by the company in December 2010, according to a letter sent to UTVIG (File: 009119 / 11-7 , 05/01/2011). According to Baxter, all customers (286) were notified and visited by company technicians for software upgrades.
  • 원인
    The equipment may issue an incorrect instruction to the user, leading to continuation of hemodialysis therapy in situations in which it should be discontinued.
  • 조치
    To health facilities: If you have this equipment at your health facility, proceed as follows: (1) Identify the software version of your equipment, following the manufacturer's instructions (see attachment); (2) If the software version of your equipment is different from versions 3.0A or 4.1B, instruct equipment operators to discontinue the treatment whenever the dialer side panel reports error codes FL04, FL07 or FL08, disregarding the UF MANUAL "instruction that will appear (this message is incorrect for the mentioned error codes). It is highly recommended that a written, easily identifiable work instruction be placed near the equipment panel that alerts operators to the problem.

Device

Manufacturer

  • Source
    ANVSANVISA