HI-TORQUE CONNECT GUIDE WIRE and HI-TORQUE CONNECT 250T GUIDE WIRE. ANVISA Register: 80146501827 and 80146501812. Models at Risk: 1012587,1012588,1012589,1012590,1012591,1012592 / 1012593, 1012594 and 1012595. All batches available. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Lake Region Medical Limited; Abbott Laboratórios do Brasil LTDA. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1333
  • 날짜
    1111-11-12
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    Abbott Vascular is initiating a voluntary Field Safety Corrective Action for the HT Connect Guide Wires peripheral wire family because a small number of devices present a partial lamination of the PTFE coating. To date, the frequency of reported incidents was 0.08%. Although no long-term or irreversible effects have been reported in patients, the potential risks associated with coating lamination include: embolism, thrombus and occlusion of the peripheral veins. Abbott Vascular research is underway and there is no definite root cause identified for inadequate adhesion of PTFE coating to some guidewires. During continued research, Abbott Vascular is identifying improvements in the coating process to ensure adhesion of PTFE coating. As the event was reported on a low frequency but randomly distributed among the manufactured batches that were distributed since the product's overseas launch in September 2011, Abbott Vascular requests that customers identify and return all HI family wires -TORQUE CONNECT GUIDE WIRE not used, regardless of batch number. #### Update (02/18/2014): According to a report sent by the company, 454 units of the product have already been collected until the date of 02/13/2014.
  • 원인
    Possibility of partial lamination of the ptfe coating, with risk of embolism, thrombi and occlusion of the peripheral veins.
  • 조치
    Abbot Vascular is collecting the product from the market. Check your inventory and segregate the identified risk products. Contact the company for return purposes.