HOMECHOICE FOR AUTOMATED PERITONEAL DIALISE. HOMECHOICE SYSTEM - VERSION 8.51 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Baxter Hospitalar Ltda 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    679
  • 날짜
    2002-10-03
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    BRAZIL - 29.10.2002 - The automated Peritoneal Dialysis System for Homechoice and Homechoice PRO is registered in Brazil under the number 10068390123. According to company information (letter dated September 27, 2002), there are no reports of physical damage associated with this situation. Baxter is sending important information to be observed with the Homechoice System, software version 8.51, launched in June 2002 to notify the clinical and patient staff of home-based DPA-directed actions to reduce the likelihood of a possible excessive infusion of volume , which could occur under certain conditions.
  • 원인
    "it was found that stopping or advancing a manual drainage during an infusion phase could result in excessive infusion of volume. therefore, if the patient requires manual drainage during the filling phase, it is important that drainage is performed without interruption, with purpose of preventing excessive volume infusion. these instructions apply only to versions 8.51 and 8.52 of the homechoice system software.
  • 조치
    Routing of safety information to customers about the Homechoice System. Proposal for corrective actions (BAXTER): - Send letters with information to affected patients and to coordinators DE Home Care; -Send revised instructions from the patient manual to Homechoice users, help reports, and an operator's manual. -Correct the possibility of excessive infusion of volume with the next version of the software. For further information please contact us on 0800 12 55 22, option 1.