MARS Treatment Kit, registration no. 80134860075. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 TERAKLIN A G 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    838
  • 날짜
    2006-09-18
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    According to Gambro Rostock, no injuries were reported in patients, however, there is a possibility of indirect loss of body fluid from the patient, caused by loss of albumin in the waste from the dialysis process. Specifically, this may be the case when the customer sets the low pressure alarm setpoint to too low relative to the current pressure too low. Inquiries or clarifications, please contact the company responsible for the product: GRIFOLS DO BRASIL LTDA (41) 3668-2444.
  • 원인
    Possibility of ruptures in the pump segment within the mars albumin circuit.
  • 조치
    Material quality controls were implemented to avoid recurrence of this type of event. In addition, the tube manufacturing process was optimized in 2005, resulting in significantly higher and longer lasting stability of the pump segment. To exclude any remote risk to patient health, in the event of pump segment ruptures, Gambro Rostock instructs all customers that "To manually change the pressure alarm set point, make sure that the difference between the current pressure and the lower alarm set point does not exceed 100 mmHg. "

Device

  • 모델명 / 제조번호(시리얼번호)
  • Manufacturer

Manufacturer

  • Source
    ANVSANVISA