OPTIUM XCEED DIABETES MONITORING SYSTEM - OPTIUM XCEED DIABETES CONTROL SYSTEM - Record no. 80146501194 - ListNo .: 70099-01, 70098-01, 70098-01. Lots: 01K180D, 01K124D, 02K102D, 02K107D, 02K127D, 03K123D, 07K199D, 05K204D, 05K204D, 06K179D, 11K179D, 07K179D, 06K179D, 06K179D, 03K201D, 06K200D, 01K202D, 08K200D, 09K179D, 06K205D, 07K123D, 08K199D, 09K123D, 09K198D E 10K177D. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Abbott Diabetes Care – ADC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    887
  • 날짜
    2007-09-17
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    Field action initiated because of Abbott Diabetes Care's ADC observation of a higher-than-expected rate of monitors exhibiting damage to the liquid crystal display (LCD) when subjected to a standard drop test. This was observed during a qualification of a new part, where part of the design check drop test involves a standard drop test. As a consequence of this observed fault of display damage, the ADC conducted the same test in the production of monitors and the same display damage was observed. However the monitors affected were only those manufactured after January 31, 2007. At the time the production monitors were identified as being affected, all monitors production and distribution were discontinued and an investigation was initiated.
  • 원인
    A higher-than-expected rate of monitors exhibiting damage to the liquid crystal display (lcd) when subjected to a standard drop test.
  • 조치
    Abbott Diabetes Care - ADC plans to insert a flyer flyer on affected lots of inventory in Brazil and also enable customers to receive the exchange of the product under warranty.