Paramount Motorized Bed. Model KA-6000. Anvisa Registry n ° 80102510312. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 VR MEDICAL IMPORTADORA E DISTRIBUIDORA DE PRODUTOS MÉDICOS LTDA.; PARAMOUNT BED CO LTD 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    1684
  • 날짜
    2015-09-16
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    According to the manufacturer of the product, the mentioned error is foreseen in product risk management and in the Equipment Instruction Manual, as well as corrective action (by the user of the product) to remedy the problem - removal of the power cable from the network for a few seconds and reconnect the device. According to the manufacturer, the error poses no risk to the safety of the patient or the operator, since it does not imply inadvertent movement of the device. After completing the investigative process in Brazil, conducted by the manufacturer of the product and by the registry holder, the root cause for the problem was the difference between the voltages in the electrical networks of the hospitals (127V) where the products are installed and the voltage of to which the product was designed (110V). This fact would be responsible for a greater occurrence of error code H01. The field action will consist of replacing part of the bed control boxes and the component called Hybrid Integrated Circuit - an integral part of the control boxes. According to the registry holder, the product at risk was distributed to only two hospitals in Brazil.
  • 원인
    Occurrence, rather than expected, of error code h01 - in which the bed controls do not respond when triggered.
  • 조치
    Field correction of physical parts of the product.