Trade name: ELEKTA PLANNING SYSTEM Technical name: 2901010 Software de Planej. of Trat. P / Radiation Therapy ANVISA Registration Number: 80569320003 Class of Risk: III Model Affected: Monaco 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 IMPAC Medical Systems, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    2069
  • 날짜
    2016-12-27
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    According to the company, it is necessary to read and return the knowledge report of the Important User Notifications to Elekta within 30 days. Request collaboration with Elekta representative to implement the solution as soon as possible. The problem does not occur every time the delete function is used in combination with the contour editing tools. Additionally, because of the complexity of the problem, it is not possible to enumerate all the workflows in which it can occur. However, you can avoid triggering the problem by embedding one of these sequences of steps into your overall workflow with the outline: 1. Delete before editing: In newly generated contours, delete any unwanted contours and save before editing (move, rotate, change format) 2. If you need to delete anything after the edits, save and download the data before deleting. This allows the affected code to find all the coordinates of the contour points. 3. Delete the outline and start over. 4. Clean the contour and start again.
  • 원인
    Changes in outline (edit and delete) no saved from program. according to the company, in a specific workflow where the contours are edited (augmented, moved, copied, etc.), then deleted in parts and then saved in the same session, it is possible that the deleted contours continue in the document, which was not to happen with the program. if a patient file is closed without saving and opened again, the bev will display, in a later created plan, the contours as if they were still there while the transverse, sagittal and coronal views will display the outlines as deleted. the total volume of the reported structure of the dvh will not coincide with structure control. if the study is saved and reopened before a new plan is created, the total volume of the structure reported in the dvh and structure control is corrected. a similar workflow can cause the defect if changes are made while the plane is loading.
  • 조치
    Field Action Code FCA-IMS-0016 triggered under the responsibility of Elekta Medical. Company will make correction in the field.