Trade name: Infusion Pump. Technical Name: Infusion Pump. ANVISA registration number: 80113010024. Risk class: III. Model Affected: Infusion Pump GH / Infusion Pump PK. Serial numbers affected: pumps manufactured prior to September / 2008 - attached list. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Respiratory Care Hospitalar LTDA.; Carefusion Switzerland 317 Sarl. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    2408
  • 날짜
    2017-11-16
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    If a pl3 alarm occurs the syringe pump should be immediately withdrawn from clinical use and should be referred for repair by qualified service personnel or authorized technical assistance. Additional information: - Date of identification of the problem by the company: 06/11/2017 - Date of notification notice to Anvisa: 11/16/2017 The company that owns the affected product is responsible for contacting its customers in due time. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • 원인
    Error code pl3 is an alarm indicating that the equipment is malfunctioning. alarm pl3 causes the device to stop working and warns the user of an internal fault by sending an error message together with a light signal and a steady high alarm that can not be silenced.
  • 조치
    Field Action Code PL3-ALARM triggered under the responsibility of Respiratory Care Hospitalar LTDA. Will make the Notice to the client.